Assembly and preparation

Assembling the breathing system

WARNING
Risk of insufficient anesthetic gas
concentrations
If the component connections of the
breathing system are not leak-tight enough,
ambient air may get into the breathing gas.
Make sure that all components of the
breathing system are connected tightly.

Preparing the ventilator

D
C
B
A
Only disinfected and sterilized components must
be used.
1 Open the ventilator door (A) with the attached
ventilator unit.
2 Release the 3 clamps (B).
3 Remove the cover (D).
62
F
G
E
4 Insert the ventilator membrane (C). After
assembly, the Dräger inscription must be
visible.
5 Fit the cover (D) and close the 3 clamps (B).
6 Connect the pressure sensor line (E) of the
ventilator chamber with the appropriate
connector.
7 Close the ventilator door with the attached
ventilator unit.
Safety functions of the ventilator
– Overpressure safety valve (F)
– Underpressure safety valve (G)
– Pressure sensor in the ventilator chamber
Mounting the CO
2
compact breathing system
WARNING
Risk of high inspiratory CO
If the soda lime is used too long, carbon
dioxide can no longer be completely
absorbed.
Check the color of the soda lime regularly,
especially if the inspiratory CO
increases unexpectedly. Replace if
necessary.
Instructions for use Fabius Tiro SW 3.n
absorber to the
values
2
value
2