Assembly and preparation
Filling and mounting
D
1 After any cleaning, push the absorber insert
back into the absorber container (D)
completely.
2 Fill CO
absorber with fresh soda lime to the
2
upper mark.
Recommendation: Only use Drägersorb 800 Plus
or Drägersorb Free.
WARNING
Risk of hypoventilation
Reuse of the soda lime dust filter can
increase filter resistance and impair the
ventilation function of Perseus.
Replace the soda lime dust filter each time
the soda lime is replaced.
E
66
F
3 Insert a new soda lime dust filter (E). Only use
soda lime dust filters indicated in the list of
accessories.
Only use undamaged filters, as exterior
damage to the filter decreases protection!
4 Insert CO
absorber (F) on the breathing
2
system from below and then rotate it counter-
clockwise as far as it will go.
Strictly observe the instructions for use for
Drägersorb 800 Plus or Drägersorb Free soda
lime.

Water trap

 Empty or replace the water trap according to its
instructions for use.
WARNING
Risk due to full water trap
When the water trap is full, the gas
measurement can fail and insufficient
ventilation may occur.
Check the water level in the water trap
regularly and empty it as necessary.
CAUTION
Risk due to gas measurement failure
Contaminants, damage, or overfilling of the water
trap can influence gas measurement.
Observe the instructions for use of the water trap.
CAUTION
Risk of misleading data
Silicon can get into the measuring cuvette and
disrupt measurement.
Do not spray the O-rings of the water trap holder
with silicon spray.
Instructions for use Perseus A500 SW 1.1n