Installing The Water Trap And Sample Line - Dräger Apollo Operating Instructions Manual

Inhalation anesthesia machine software 3.2n
Hide thumbs Also See for Apollo:
Table of Contents

Advertisement

Installing the Water Trap and Sample Line

1. Connect 22 mm (0.87 in) breathing hoses to the
inspiratory and expiratory ports on the breathing
system (1 in
Figure
2. Connect the other end of the breathing hoses to
a Y-piece (2 in
Figure
on the Y-piece.
Installing the Water Trap and
Sample Line
1. Push the new or empty water trap into its holder
on the front of the machine(3 in
clicks into place.
2. Connect one end of the sample line to the Luer
fitting on the water trap (4 in
3. Connect the other end of the sample line to the
Luer fitting on the Y-piece (2 in
Ensure that all Luer fittings are securely
connected.
Note:
Use only original sample line. Other lines
may change the technical data of the device.
Note:
For sample lines available for use with the
Apollo, see the Accessories List, P/N
8603528.
CAUTION !
Sample gas lines are single-use articles and must
not be disinfected.
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
WARNING !
The water trap must be replaced after a service
life of four weeks.
The diaphragm could become brittle after an
extended period of use, allowing water and
bacteria to get into the measurement system.
The contamination affects the gas
measurement, or the gas measurement may fail
as a result.
Apollo Operator's Manual
33).
33), or to the optional filter
Figure
33) until it
Figure
33).
Figure
33).
Chapter 4 - System Setup
Figure 33. Breathing Hose and Water Trap Connections
1
3
4
1
2
51