Connection of the pressure gauge for
measurement of the airway pressure
(optional)
C
D
B
1 Connect the pressure measurement hose to
the corresponding connector (A) on the bottom
of the compact breathing system.
2 Connect the T-piece (B) with the pressure
measurement hose and the pressure gauge.
3 Connect the other end of the pressure
measurement hose to the bacterial filter (C).
4 Plug in the bacterial filter into the connection
(D) marked with
Instructions for use Fabius Tiro SW 3.n
A
on the rear of the device.
Connecting the flow sensor
1 Connect the flow sensor cable to the
corresponding connection (A) on the bottom of
the compact breathing system.
2 Plug in the other end of the flow sensor cable
into the connection (B) marked with
rear of the device.
Assembly and preparation
B
A
on the
77