Connection Of The Pressure Gauge For Measurement Of The Airway Pressure (Optional); Connecting The Flow Sensor - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Assembly and preparation
Connection of the pressure gauge for
measurement of the airway pressure
(optional)
D
1 Connect the pressure measurement hose (C)
to the corresponding connection on the bottom
of the compact breathing system.
2 Connect the T-piece (B) with the pressure
measurement hose and the pressure gauge.
3 Connect the other end of the pressure
measurement hose with the bacterial filter (B).
4 Plug in the bacterial filter into the connection
(D) marked with
80
C
B
on the rear of the device.

Connecting the flow sensor

A
1 Connect the flow sensor cable to the
corresponding connection (A) on the bottom of
the compact breathing system.
2 Plug in the other end of the flow sensor cable
into the connection (B) marked with
rear of the device.
Instructions for use Fabius plus SW 3.n
B
A
on the