Waters Autospec Premier Operator's Manual page 73

Table of Contents

Advertisement

Magnet Scan function editor parameters (Continued)
Parameter
Inter-Scan Delay (s)
Total Cycle Time (s)
To set up a Magnet Scan function
1.
In the MS method editor, click
Click Functions > Magnet Scan.
Alternative:
2.
In the Magnet Scan function editor, enter the required parameters.
3.
Click OK.
See
"The Magnet Constant Utility dialog box" on page
Description
The time between one scan's end and the next scan's
start, in seconds. No data are collected during this
time; the time is used to allow the magnet to settle,
and to transfer collected data from the EPC to the
MassLynx PC.
This value is typically set to 10% of the scan
Tip:
time (s per decade). However, if the calibration at
the high mass end is poor, the value set may be too
small.
Displays the total time per scan cycle, in seconds.
.
D-14.
Setting up scanning functions
3-9