EI sensitivity chromatogram
14. Repeat
steps 7 and 10
To process the data
1.
In the chromatogram window, click Edit > Integrated Peaks.
2.
Zoom in on the first desorption curve.
3.
Right-click and drag along the horizontal axis to define the area of the
curve over which to integrate.
4.
In the Edit Integrated Peaks dialog box, click Add.
5.
Record this value of the area (A1).
6.
Repeat
steps 2 to 4
and A3).
7.
Calculate the average area of the peaks, using the values in the
following figure as an example:
A1
A2
A3
+
+
A
=
---------------------------------- -
3
Where A is the average area of the peaks in bit-minutes.
twice, using a new sample cup each time.
for the remaining two desorption curves (Areas A2
7232969
+
7400556
+
7963153
=
---------------------------------------------------------------------------
3
Measuring EI sensitivity from the solids probe
7-87