AutoSpec Tune Acquisition dialog box parameters (Continued)
Parameter
Data File Name
Text
Origin
Acquisition Function Specifies how data are collected (select one):
Data Format
Scan per spectrum
Low Mass (m/z)
High Mass (m/z)
Resolution
Duration (mins)
Scan Time (s)
Description
Specifies the name of the data file to be created in
this text box.
Provide a brief description of the sample. The
description can be displayed on any output of the
acquired data.
Opens the Sample Origin dialog box. You can then
specify additional information about the sample in
the following fields:
• Submitter
• Job
• Task
• Conditions
• Magnet scan
• Voltage scan
• B/E scan
2
• B
/E scan
• CNL scan
If you select B/E scan, B
Note:
the Linked Scanning Parameters options are
enabled.
Specifies the type of data (Centroid or Continuum)
to be collected and stored on disk.
If B/E scan, B
Note:
selected, only Continuum is available.
Displays the number of scans per spectrum display.
The mass at which the function starts.
The mass at which the function stops.
The resolution, used to determine scanning
parameters.
The length of the acquisition, in minutes.
The length of each scan, in seconds.
2
/E scan, or CNL scan,
2
/E scan, or CNL scan is

Starting data acquisition

4-3