Waters Autospec Premier Operator's Manual page 162

Table of Contents

Advertisement

GC interface
To renew the injection port septum
Warning:
with toxic materials. Wear nitrile gloves at all times while
handling the components.
Warning:
interface, because it can be hot.
1.
Open the reference inlet and pump valves.
2.
Wait for 10 minutes to remove any sample from the injection port.
3.
Close the reference inlet valve.
4.
On the MassLynx Tune window's Inlets tab, set Capillary Line 1,
Capillary Line 2, and Re-entrant to 50, and Septum to 0.
5.
Wait for approximately 45 minutes for the GC interface to cool.
Monitor the Septum readback to ensure that the temperature is less
Tip:
than 40 °C.
6.
Unscrew and remove the injection port thumbscrew.
7-4
Routine maintenance
Reference inlet valve
Pump valve
The GC interface components can be contaminated
To avoid burns, take care when working with the GC
Injection port
Thumbscrew