This appendix details the performance specifications, and the types of
file used by MassLynx software, for the instrument.
Contents
Topic
Performance specifications
File types
Performance specifications
The following specifications are for the Ultima (standard) magnet.
Note:
Contact Waters for specification details if your instrument is fitted with the
optional High-field or Low-field magnets.
Ultimate Resolving Power (RP): 80,000
Transmission at 10,000 RP relative to 1,000 RP: 10%
Ultimate sensitivity probe, EI, methyl stearate, 1,000 RP:
-7
5 × 10
GC/MS sensitivity for 2,3,7,8-TCDD:
In the SIR mode at 10,000 resolution, while selecting two masses from
natural 2,3,7,8-TCDD (m/z 319.8965 and 321.8936) and two masses from
13
C
-2,3,7,8-TCDD (m/z 331.9368 and 333.9339) plus a lock-mass and
12
lock-mass check; the signal-to-noise ratio (S/N) at m/z 321.8936 from
100 fg of 2,3,7,8-TCDD standard injected on to a 30 metre DB5 column is
as follows:
S/N
encompassing 95% of raw noise data points (equivalent to 4
standard deviations, that is. ±2σ about the mean).
coulombs/µg
125:1, where noise is a peak-to-peak measurement