Resolution Checker dialog box parameters (Continued)
Parameter
All Masses
High and Low
Masses
Print Resolution
Check
Use Function
One Only
To open an existing resolution check setup file
1.
Click File > Open.
2.
Enter the required experiment file name, or select an existing file from
the displayed list.
3.
Click OK.
To save the current resolution check setup file
Click File > Save.
To save the current resolution check setup file to a new file name:
1.
In the Resolution Checker dialog box, click Click File > Save As.
2.
Enter a new file name, or select an existing file from the displayed list.
3.
Click Save.
To verify VSIR experimental resolution
1.
In the MassLynx sample list, in the Process column, select Resolution
Check.
2.
In the Process Options column, enter the full path and file name of the
resolution check setup file, for example,
c:\masslynx\autospec\resolution.dat.
3.
In the MassLynx Tune window, click Options > Resolution Checker.
Description
Specifies that the resolution check is performed on all
the masses in the VSIR function mass range.
Specifies that the resolution check is performed on
only the highest and lowest masses in the VSIR
function mass range.
Prints a report on completion of the resolution check.
Specifies that the resolution check is performed only
on the first function of the experiment.
The Tune window Options menu
D-25