4.
Fit a ceramic spacer, with the stepped portion uppermost, over each of
the locating dowels.
You must fit the bayonet support so that the connector slit is
Caution:
parallel to the repeller contact.
5.
Fit the bayonet support to the ceramic spacers.
Connector slit
Repeller contact
Source maintenance