AutoSpec Tune Acquisition dialog box parameters (Continued)
Parameter
Inter Scan Delay (s)
Linked scanning options are available only if you select B/E scan,
Note:
2
B
/E scan, or CNL scan in the Acquisition Function list box.
Function Mass
Low Mass Precursor
Ion (m/z)
High Mass Precursor
Ion (m/z)
Pre-Acquisition
Calibration File
Name
Start
Close
Calibration
4-4
Acquiring data
Description
The time between one scan's end and the next scan's
start, in seconds. No data are stored in this period.
This value is typically 10% of Scan Time.
Tip:
Enter the required mass appropriate to the current
acquisition function:
• For a B/E Scan, the precursor mass.
2
• For a B
/E Scan, the product mass.
• For a CNL Scan, the loss mass.
Display only.
Display only.
Specify a pre-acquisition calibration file name (of
type *.cal, or *.scl) and path in the text box, or click
Browse.
This file is used to calibrate the scan of ESA1
Rule:
and ESA2 performed by the current Acquisition
Function scan. This file must be created on standard
Magnet Scan data, over exactly the same magnet
mass range and scan time as the scan.
Starts the acquisition.
Closes this dialog box. The current settings are
retained when the dialog box reopens.
Opens the Scope Calibration dialog box, from which
you to select a calibration file to apply to the
acquired data.