Sweep Mode; Sweep Time - Keysight Technologies 33210A User Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

3
Features and Functions

Sweep Mode

You can sweep with either linear or logarithmic spacing. For a linear sweep, the
function generator varies the output frequency in a linear fashion during the
sweep. For a logarithmic sweep, the function generator varies the output
frequency in a logarithmic fashion.
– Sweep mode: Linear or Logarithmic. The default is Linear.
– Front-Panel Operation: After enabling sweeps, press the Linear softkey again
– Remote Interface Operation:

Sweep Time

The sweep time specifies the number of seconds required to sweep from the start
frequency to the stop frequency. The number of discrete frequency points in the
sweep is automatically calculated by the function generator and is based on
the sweep time you select.
– Sweep time: 1 ms to 500 seconds. The default is 1 second.
– Front-Panel Operation: After enabling sweeps, press the Sweep Time softkey.
– Remote Interface Operation:
108
to toggle between the linear or log mode.
SWEep:SPACing {LINear|LOGarithmic}
Then use the knob or numeric keypad to enter the desired sweep time.
SWEep:TIME {<seconds>|MINimum|MAXimum}
Keysight 33210A User's Guide