Keysight Technologies 33210A User Manual page 201

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

VOLTage:UNIT {VPP|VRMS|DBM}
VOLTage:UNIT?
Select the units for output amplitude (does not affect offset voltage or high/low
levels). The default is VPP. The :UNIT? query returns "VPP", "VRMS", or "DBM".
– The function generator uses the current units selection for both front panel and
– The VOLT? query command (see page 192) returns the output amplitude in the
– The output units for amplitude cannot be set to dBm if the output termination
– Unless you specify the units as part of the APPLy or VOLT command,
Keysight 33210A User's Guide
remote interface operations. For example, if you select "VRMS" from the
remote interface using the VOLT:UNIT command, the units are displayed as
"VRMS" on the front panel.
units set by the most recent VOLT:UNIT command.
is currently set to "high impedance". The units are automatically converted to
Vpp. See the OUTP:LOAD command on page 199 for more information.
the VOLT:UNIT command takes precedence. For example, if you select "Vrms"
using the VOLT:UNIT command and do not include units with the APPLy or
VOLT command, the value specified for the amplitude parameter in the APPLy
command will be in "Vrms".
Remote Interface Reference
4
201