Keysight Technologies 33210A User Manual page 156

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

3
Features and Functions
To Unsecure for Calibration
You can unsecure the function generator either from the front panel or over the
remote interface. The function generator is secured when shipped from the
factory and the security code is set to "AT33210A".
– Once you enter a security code, that code must be used for both front-panel
– Front-Panel Operation: Press
– Remote Interface Operation: To unsecure the function generator, send the
To Secure Against Calibration
You can secure the function generator either from the front panel or over the
remote interface. The function generator is secured when shipped from the factory
and the security code is set to "AT33210A".
– Once you enter a security code, that code must be used for both front-panel
– Front-Panel Operation: Press
– Remote Interface Operation: To secure the function generator, send the
156
and remote operation. For example, if you secure the function generator from
the front panel, you must use that same code to unsecure it from the remote
interface.
from the "Test / Cal" menu.
following command with the correct security code.
CAL:SECURE:STATE OFF,AT33210A
and remote operation. For example, if you secure the function generator from
the front panel, you must use that same code to unsecure it from the remote
interface.
from the "Test / Cal" menu.
following command with the correct security code.
CAL:SECURE:STATE ON,AT33210A
and then select the Secure Off softkey
and then select the Secure On softkey
Keysight 33210A User's Guide