Device Dependent Errors

-313
Calibration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator's calibration
constants has detected a checksum error. This error may be the result of a device
failure or extreme conditions such as lightning or strong magnetic fields.
-314
Save/recall memory lost;
memory corruption detected
The non-volatile memory used to store instrument states has detected a
checksum error. This error may be the result of a device failure or extreme
conditions such as lightning or strong magnetic fields.
-315
Configuration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator's configuration
settings (i.e., remote interface settings) has detected a checksum error. This error
may be the result of a device failure or extreme conditions such as lightning or
strong magnetic fields.
-350
Queue overflow
The error queue is full because more than 20 errors have occurred. No additional
errors are stored until you remove errors from the queue. The error queue is
cleared by the *CLS (clear status) command or when the power is cycled. The
errors are also cleared when you read the queue. The error queue is not cleared by
an instrument reset (*RST command).
318
Keysight 33210A User's Guide