Burst Mode Commands - Keysight Technologies 33210A User Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

Burst Mode Commands

Use the APPLy command or the equivalent FUNC, FREQ, VOLT, and VOLT:OFFS
NOTE
commands to configure the waveform. For internally-triggered bursts, the
minimum frequency is 2.001 mHz. For sine and square waveforms, frequencies
above 3 MHz are allowed only with an "infinite" burst count.
BURSt:MODE {TRIGgered|GATed}
BURSt:MODE?
Select the burst mode. In the triggered mode, the function generator outputs a
waveform with a specified number of cycles (burst count) each time a trigger is
received from the specified trigger source (TRIG:SOUR command). In the gated
mode, the output waveform is either "on" or "off " based on the level of the
external signal applied to the rear-panel Trig In connector. The default is TRIG.
The :MODE? query returns "TRIG" or "GAT".
– When the gated mode is selected, the waveform generator either runs or stops
– When the gated mode is selected, the burst count, burst period, and trigger
Keysight 33210A User's Guide
based on the logic level of the gate signal applied to the rear-panel Trig In
connector. You can select the polarity of the Trig In connector using the
BURS:GATE:POL command (see page 236). When the gate signal is true, the
function generator outputs a continuous waveform. When the gate signal goes
false, the current waveform cycle is completed and then the function
generator stops while remaining at the voltage level corresponding to the
starting burst phase of the selected waveform.
source are ignored (these parameters are used for the triggered burst mode
only). If a manual trigger is received (TRIG command), it is ignored and no
error will be generated.