-222
Data out of range;
pulse d uty cycle limited by period; value clipped to ...
The specified pulse duty cycle must conform to the following period and edge
time restriction. The function generator will adjust the pulse duty cycle as needed
to accommodate the specified period.
-222
Data out of range;
large period limits minimum pulse wid th
For a pulse waveform, the minimum pulse width is 40 ns for periods up to 10
seconds. For periods above 10 seconds, the minimum pulse width is greater. See
Pulse Configuration Commands
-222
Data out of range;
pulse edge time limited by wid th; value clipped to ...
The specified edge time must fit within the specified pulse width as shown below.
The function generator will adjust the edge time as needed to accommodate the
specified pulse width.
-222
Data out of range;
pulse edge time limited by d uty cycle; value clipped to ...
The specified edge time must fit within the specified pulse duty cycle as shown
below. The function generator will adjust the edge time as needed to
accommodate the specified pulse duty cycle.
Keysight 33210A User's Guide
Duty Cycle < 100 × (Period – (1.6 × Edge Time)) ÷ Period
Edge Time < 0.625 × Pulse Width
Edge Time < 0.625 × Period × Duty Cycle ÷ 100