TRIGger:SLOPe {POSitive|NEGative}
TRIGger:SLOPe?
Select whether the function generator uses the rising edge or falling edge of the
trigger signal on the rear-panel Trig In connector for an externally-triggered
sweep. The default is POS (rising edge). The :SLOP? query returns "POS" or
"NEG".
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger:SLOPe?
Select a rising or falling edge for the "trigger out" signal. When enabled using the
OUTP:TRIG command (see below), a TTL-compatible square waveform with the
specified edge is output from the rear-panel Trig Out connector at the beginning
of the sweep. Select "POS" to output a pulse with a rising edge or select "NEG" to
output a pulse with a falling edge. The default is POS. The :SLOP? query returns
"POS" or "NEG".
– When the Immediate (internal) trigger source is selected (TRIG:SOUR IMM
– When the External trigger source is selected (TRIG:SOUR EXT command), the
– When the Bus (software) trigger source is selected (TRIG:SOUR BUS
OUTPut:TRIGger {OFF|ON}
OUTPut:TRIGger?
Disable or enable the "trigger out" signal. When enabled, a TTL-compatible
square waveform with the specified edge (OUTP:TRIG:SLOP command) is output
from the rear-panel Trig Out connector at the beginning of the sweep. The default
is OFF. The :TRIG? query returns "0" (OFF) or "1" (ON).
Keysight 33210A User's Guide
command), the function generator outputs a square waveform with a 50%
duty cycle (the rising edge is the sweep trigger) from the Trig Out connector.
The period of the waveform is equal to the specified sweep time (SWE:TIME
command).
function generator automatically disables the "trigger out" signal. The
rear-panel Trig Out connector cannot be used for both operations at the same
time (an externally-triggered sweep uses the same connector to trigger the
sweep).
command), the function generator outputs a pulse (>1 µs pulse width) from
the Trig Out connector at the beginning of each sweep.