Output Termination - Keysight Technologies 33210A User Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

3
Features and Functions

Output Termination

Applies to output amplitude and offset voltage only. The Keysight 33210A has a
fixed series output impedance of 50 ohms to the front-panel Output connector. If
the actual load impedance is different than the value specified, the displayed
amplitude and offset levels will be incorrect.
– Output termination: 1Ω to 10 kΩ, or Infinite. The default is 50Ω. The message
– The output termination setting is stored in non-volatile memory and does not
– If you specify a 50-ohm termination but are actually terminating into an open
– If you change the output termination setting, the displayed output amplitude
– You cannot specify the output amplitude in dBm if the output termination is
– Front-Panel Operation: Press
– Remote Interface Operation:
76
line at the top of the display calls attention to output termination settings
other than 50Ω.
change when power has been off or after a remote interface reset (assuming
the Power On state is set to "default").
circuit, the actual output will be twice the value specified. For example, if you
set the offset to 100 mVdc (and specify a 50-ohm load) but are terminating the
output into an open circuit, the actual offset will be 200 mVdc.
and offset levels are automatically adjusted (no error will be generated). For
example, if you set the amplitude to 10 Vpp and then change the output
termination from 50 ohms to "high impedance", the amplitude displayed on
the function generator's front-panel will double to 20 Vpp. If you change from
"high impedance" to 50 ohms, the displayed amplitude will drop in half.
currently set to "high impedance". The units are automatically converted to
Vpp.
use the knob or numeric keypad to select the desired load impedance or press
the Load softkey again to choose "High Z".
OUTPut:LOAD {<ohms>|INFinity|MINimum|MAXimum}
and select the Output Setup softkey. Then
Keysight 33210A User's Guide