2. EXT 1 & EXT 2
Connector
female BNC
Signal
An externally supplied ±1 V
Damage Levels
5 V
and 10 V
rms
3. LF OUT
Connector
female BNC
Signal
Voltage range: 0 to +5 V
Offset: -5 V to +5 V, nominal
For more information, see page
4. SWEEP OUT
Connector
female BNC
Can drive 2 kΩ.
Signal
Voltage range: 0 to +10 V, regardless of sweep width
In swept mode: beginning of sweep = 0 V; end of sweep = +10 V
In CW mode: no output
This is a multiple use connector. For signal routing selections, see pages
5. PULSE
Connector
female BNC
Signal
Externally supplied: +1 V = on; 0 V = off
Damage Levels
≤ −0.3 and ≥+5.3 V
6. TRIG 1 & 2
Connector
female BNC
Signal
An externally supplied TTL or CMOS signal for triggering operations, such as point to point in
manual sweep mode or an LF sweep in external sweep mode.
Triggering can occur on either the positive or negative edge.
Damage Levels
≤ −0.5 and ≥+5.5 V
7. REF IN
Connector
female BNC
Signal
An externally supplied −3.5 to +20 dBm signal from a timebase reference that is within ±1 ppm.
In its factory default mode, the signal generator can detect a valid reference signal at this
connector and automatically switch from internal to external reference operation. See
the Signal Generator" on page
tell the signal generator the external reference frequency you wish to use; enter the information
through the front panel or over the remote interface.
Keysight EXG and MXG X-Series Signal Generators User's Guide
Rear Panel Overview (N5171B, N5172B, N5181B, & N5182B)
Impedance nominally 50 Ω
signal that produces the indicated depth.
p
p
Impedance 50 Ω
p
.
165
Impedance <1 Ω
Impedance nominally 50 Ω
Impedance high Z
Impedance nominally 50 Ω
40. With Option 1ER (flexible reference input), you must explicitly
Signal Generator Overview
and
.
47
169
"Presetting
13