Table of Contents

Advertisement

4
Remote Interface Reference
4 Set the sweep time.
5 Select the sweep trigger source.
6 Set the marker frequency. (Optional)
7 Enable the sweep mode.
222
Use the SWE:TIME command to set the number of seconds required to sweep
from the start frequency to the stop frequency.
Use the TRIG:SOUR command to select the source from which the sweep will
be triggered.
If desired, you can set the frequency at which the signal on the front-panel
Sync connector goes to a logic low during the sweep. Use the MARK:FREQ
command to set the marker frequency to any value between the start
frequency and stop frequency. Use the MARK ON command to enable the
frequency marker.
After you have set up the other sweep parameters, use the SWE:STAT ON
command to enable the sweep mode.
Keysight 33210A User's Guide