Remote Interface Reference

Pulse Configuration Commands

See also
This section describes the low-level commands used to program the function
generator to output a pulse waveform. To select the pulse function, use the
FUNC PULS command (see page 189). Refer to the diagram below for the
command descriptions that follow.
50%
PULSe:PERiod {<seconds>|MINimum|MAXimum}
PULSe:PERiod? [MINimum|MAXimum]
Set the period for pulses. Select a period from 200 ns to 1000 seconds.
The default is 1 ms. MIN = 200 ns. MAX = 1000 seconds. The PULS:PER? query
returns the period of the pulse waveform in seconds.
– The specified period must be greater than the sum of the pulse width and the
202
"Pulse Waveforms"
90%
Pulse Width
10%
Rise Time
edge time. The function generator will adjust the edge time and the pulse
width as needed to accommodate the specified period. From the remote
interface, a "Settings conflict" error will be generated. The edge time is
minimized first, and then the width (or duty cycle) is adjusted as shown below.
Period > Pulse Width + (1.6 × Edge Time)
or in terms of pulse duty cycle:
Period > (Period × Duty Cycle ÷ 100) + (1.6 × Edge Time)