10 Define the remaining waveform points.
Keysight 33210A User's Guide
Using the Time and Voltage softkeys, define the remaining waveform points
using the values shown in the table below.
Point
1
2
3
4
– The time value of the last point that can be defined in the waveform must
be less than the specified cycle period.
– The waveform editor automatically connects the last waveform point to the
voltage level of first point to create a continuous waveform.
– To insert additional points after the current waveform point, press the
Insert Point softkey. The new point is inserted midway between the current
point and the next defined point.
– To remove the current waveform point, press the Remove Point softkey.
The remaining points are joined using the interpolation method currently
selected. You cannot remove Point #1 since the waveform must have a
defined initial value.
Features and Functions
Time Value
0 s
2 ms
4 ms
7 ms
3
Vol tage Value
0 V
3 V
1 V
0 V