Modulating Source

The function generator will accept an internal or external modulation source for
PWM.
– Modulating source: Internal or External. The default is Internal.
– If you select the External source, the pulse waveform is modulated with an
– Front-Panel Operation: After enabling PWM, press the Source softkey.
– Remote Interface Operation:
104
external waveform. The width deviation or duty cycle deviation is controlled by
the ±5V signal level present on the rear-panel Modulation In connector.
For example, if you have set the duty cycle to 10% and the duty cycle deviation
to 5%, then when the modulating signal is at +5 volts, the output will be at the
maximum duty cycle (15%). When the modulating signal is at -5 volts, the output
will be at the minimum duty cycle (5%).
Modulation
In
PWM:SOURce {INTernal|EXTernal}
+5 V
0 V
-5 V