3
Features and Functions

Pulse Width

The pulse width represents the time from the 50% threshold of the rising edge of
the pulse to the 50% threshold of the next falling edge.
– Pulse width: 40 ns to 1000 s (see restrictions below). The default pulse width is
– The minimum pulse width (Wmin) is affected by the period.
– The specified pulse width must also be less than the difference between the
– The specified pulse width must be less than the difference between the period
– The pulse width must also be greater than the total time of one edge as shown
– Front-Panel Operation: After selecting the pulse function, press the Wid th
– Remote Interface Operation:
84
100 µs.
Wmin = 40 ns for period < 10 s.
Wmin = 200 ns for period > 10 s, but < 100 s.
Wmin = 2 µs for period > 100 s, but < 1000 s.
period and the minimum pulse width as shown below. The function generator
will adjust the pulse width as needed to accommodate the specified period.
Pulse Width < Period – Wmin
and the edge time as shown below. The function generator will automatically
adjust the pulse width as needed to accommodate the specified period.
Pulse Width < Period – (1.6 × Edge Time)
below.
Pulse Width > 1.6 × Edge Time
softkey. Then use the knob or numeric keypad to enter the desired pulse width.
FUNCtion:PULSe:WIDTh {<seconds>|MINimum|MAXimum}
Keysight 33210A User's Guide

Advertisement

Table of Contents
loading

Table of Contents