Error Messages

Calibration Errors

The following errors indicate failures that may occur during a calibration procedure
(see Chapter 4 in the Keysight 33210A Service Guide).
701
Calibration error; security defeated by hard ware jumper
The function generator's calibration security feature has been disabled by
temporarily shorting the two "CAL ENABLE" pads on the internal circuit board, as
described in the Keysight 33210A Service Guide.
702
Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured. To
unsecure the instrument, use the CAL:SEC:STAT ON command with the correct
security code.
703
Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command.
706
Calibration error; provided value is out of range
The calibration value specified with the CAL:VAL command is out of range.
707
Calibration error; signal input is out of range
The internal analog-to-digital converter (ADC) has determined that the signal
applied to the rear-panel Modulation In connector is out of range.