Error Messages
5
-222
Data out of range;
pulse wid th; value clipped to ...
This generic message indicates that the desired pulse width is limited to an upper
or lower boundary which is dictated by the instrument hardware.
-222
Data out of range;
pulse edge time; value clipped to ...
This generic message indicates that the desired edge time is limited to an upper
or lower boundary which is dictated by the instrument hardware.
-222
Data out of range;
FM deviation; value clipped to ...
This generic message indicates that the desired FM deviation is limited to a lower
or upper boundary as set by the current function's frequency.
-222
Data out of range;
FM deviation limited by minimum frequency
The frequency deviation is limited to the lower limit (1 mHz).
-222
Data out of range;
FM deviation limited by maximum frequency; value clipped to upper limit
The frequency deviation cannot exceed the carrier frequency, and is limited as
follows: 5 MHz for sine or square wave carriers.