Frequency Sweep Commands

See also

Sweep Overview

The following is an overview of the steps required to generate a sweep. The
commands used for sweep are listed on page 223.
1 Select the waveform shape, amplitude, and offset.
2 Select the frequency boundaries of the sweep.
3 Select the sweep mode.
Keysight 33210A User's Guide
"Frequency Sweep"
Use the APPLy command or the equivalent FUNC, FREQ, VOLT, and VOLT:OFFS
commands to select the function, frequency, amplitude, and offset. You can
select a sine, square, or ramp waveform (pulse, noise, dc, and arbitrary are not
allowed).
You can set the frequency boundaries using one of two methods.
a Start Frequency/Stop Frequency: Use the FREQ:STAR command to set the
start frequency and the FREQ:STOP command to set the stop frequency.
To sweep up in frequency, set the start frequency < stop frequency.
To sweep down in frequency, set the start frequency > stop frequency.
b Center Frequency/Frequency Span: Use the FREQ:CENT command to set
the center frequency and the FREQ:SPAN command to set the frequency
span.
To sweep up in frequency, set a positive span.
To sweep down in frequency, set a negative span.
Use the SWE:SPAC command to select linear or logarithmic spacing for the
sweep.