5
Error Messages
-221
Settings conflict;
high level changed d ue to low level
You can set the levels to a positive or negative value, but note that the high level
must always be greater than the low level. If you specify a low level that is greater
than the existing high level, the function generator will automatically set the high
level to be 1 mV greater than the low level.
-222
Data out of range;
value clipped to upper limit
The specified parameter is outside of the function generator's capability.
The function generator has adjusted the parameter to the maximum value
allowed. Example: PHAS 1000
-222
Data out of range;
value clipped to lower limit
The specified parameter is outside of the function generator's capability.
The function generator has adjusted the parameter to the minimum value
allowed. Example: PHAS -1000
-222
Data out of range;
pulse edge time limited by period; value clipped to upper limit
The specified edge time must fit within the existing period and width. The function
generator will limit the edge time as needed to accommodate the existing period.
-222
Data out of range;
pulse wid th limited by period; value clipped to ...
The specified pulse width must be less than the difference between the period
and the edge time as shown below. The function generator will adjust the pulse
width as needed to accommodate the specified period.
310
Pulse Width < Period – (1.6 × Edge Time)