Keysight Technologies 33210A User Manual page 219

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

– The duty cycle deviation is also limited by the minimum pulse width (Wmin):
– The duty cycle deviation is also limited by the current edge time setting.
– If you select the External modulating source (PWM:SOUR EXT command), the
The operation of the PWM:DEV:DCYC command is affected by the
NOTE
FUNC:PULS:HOLD command (see
for further information). The FUNC:PULS:HOLD command determines whether
pulse width (the default) or pulse duty cycle values are to be held constant as
the period is varied. If the width is held, so is the width deviation. If the duty
cycle is held, so is the duty cycle deviation. If pulse wid th and width deviation
are being held, duty cycle deviation values specified with the PWM:DEV:DCYC
command are automatically converted to the equivalent wid th deviation in
seconds.
Keysight 33210A User's Guide
Duty Cycle Deviation < Duty Cycle – 100 × Wmin ÷ Period
and
Duty Cycle Deviation < 100 – Duty Cycle – 100 × Wmin ÷ Period
where:
Wmin = 40 ns for period < 10 s.
Wmin = 200 ns for period > 10 s, but < 100 s.
Wmin = 2 µs for period > 100 s, but < 1000 s.
Duty Cycle Dev < Duty Cycle – (160 × Edge Time) ÷ Period
and
Duty Cycle Dev < 100 – Duty Cycle – (160 × Edge Time) ÷ Period
deviation is controlled by the ±5V signal level present on the rear-panel
Modulation In connector. For example, if you have set the duty cycle deviation
to 5 percent, a +5V signal level corresponds to a 5 percent deviation, that is an
additional 5% of period added to the pulse duty cycle. Lower external signal
levels produce less deviation, and negative signal levels reduce the duty cycle.
Remote Interface Reference
"Pulse Configuration
Commands" on page 202
4
219

Advertisement

Table of Contents
loading

Table of Contents