5
Error Messages
-221
Settings conflict;
trigger output connector used by burst gate
If you have selected the gated burst mode (BURS:MODE GAT command) with burst
enabled, the "trigger out" signal cannot be enabled (OUTP:TRIG ON command).
The rear-panel Trig connector cannot be used for both operations at the same
time.
-221
Settings conflict;
trigger output connector used by trigger external
When the external trigger source is selected (TRIG:SOUR EXT command),
the function generator automatically disables the "trigger out" signal.
The rear-panel Trig connector cannot be used for both operations at the same
time.
-221
Settings conflict;
frequency red uced for user function
For arbitrary waveforms, the output frequency is limited to 3 MHz. When you
change to the arbitrary waveform function (APPL:USER or FUNC:USER command)
from a function that allows a higher frequency, the function generator will
automatically adjust the frequency to 3 MHz.
-221
Settings conflict;
frequency changed for pulse function
For pulse waveforms, the output frequency is limited to 5 MHz. When you change
to the pulse function (APPL:PULS or FUNC:PULS command) from a function that
allows a higher frequency, the function generator will automatically adjust the
frequency to 5 MHz.
300
Keysight 33210A User's Guide