– Function Limitations: If you change to a function whose maximum frequency is
– Amplitude Limitations: If you change to a function whose maximum amplitude
– Front-Panel Operation: To select a function, press any key in the top row of
– Remote Interface Operation:
Keysight 33210A User's Guide
less than that of the current function, the frequency is adjusted to the
maximum value for the new function. For example, if you are currently
outputting a 10 MHz sine wave and then change to the ramp function, the
function generator will automatically adjust the output frequency to 100 kHz
(the upper limit for ramps).
is less than that of the current function, the amplitude is automatically
adjusted to the maximum value for the new function. This may occur when the
output units are Vrms or dBm due to the differences in crest factor for the
various output functions.
For example, if you output a 5 Vrms square wave (into 50 ohms) and then
change to the sine wave function, the function generator will automatically
adjust the output amplitude to 3.536 Vrms (the upper limit for sine in Vrms).
function keys. Press
To view the other arbitrary waveform choices, press the Select Wform softkey.
To select dc volts from the front panel, press
softkey. Press the Offset softkey to enter the desired offset voltage level.
FUNCtion {SINusoid|SQUare|RAMP|PULSe|NOISe|DC|USER}
You can also use the APPLy command to select the function, frequency,
amplitude, and offset with a single command.
to output the arbitrary waveform currently selected.
Features and Functions
and then select the DC On
3
69