Keysight Technologies 33210A User Manual page 186

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

4
Remote Interface Reference
APPLy:SINusoid [<frequency> [,<amplitude> [,<offset>] ]]
Output a sine wave with the specified frequency, amplitude, and dc offset. The
waveform is output as soon as the command is executed.
APPLy:SQUare [<frequency> [,<amplitude> [,<offset>] ]]
Output a square wave with the specified frequency, amplitude, and dc offset. This
command overrides the current duty cycle setting and automatically selects 50%.
The waveform is output as soon as the command is executed.
APPLy:RAMP [<frequency> [,<amplitude> [,<offset>] ]]
Output a ramp wave with the specified frequency, amplitude, and dc offset. This
command overrides the current symmetry setting and automatically selects
100%. The waveform is output as soon as the command is executed.
APPLy:PULSe [<frequency> [,<amplitude> [,<offset>] ]]
Output a pulse wave with the specified frequency, amplitude, and dc offset. The
waveform is output as soon as the command is executed.
– This command preserves the current pulse width setting (FUNC:PULS:WIDT
186
command) or pulse duty cycle setting (FUNC:PULS:DCYC command),
depending on which has been selected to "hold" (FUNC:PULS:HOLD
command). The edge time setting (FUNC:PULS:TRAN command) is also
preserved. However, based on the specified frequency, the function generator
will adjust the pulse width or edge time to comply with the frequency
restrictions for pulse waveforms. See page 202 for more information on setting
the pulse width and edge time.
Keysight 33210A User's Guide

Advertisement

Table of Contents
loading

Table of Contents