Keysight Technologies 33210A User Manual page 235

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

– When the Bus (software) source is selected, the function generator outputs
– When the External or Bus trigger source is selected, the burst count and burst
– The APPLy command automatically sets the trigger source to Immediate
– To ensure synchronization when the Bus source is selected, send the *WAI
– You can use the *OPC? (operation complete query) command or the *OPC
TRIGger:SLOPe {POSitive|NEGative}
TRIGger:SLOPe?
Select whether the function generator uses the rising edge or falling edge of the
trigger signal on the rear-panel Trig In connector for an externally-triggered burst.
The default is POS (rising edge). The :SLOP? query returns "POS" or "NEG".
Keysight 33210A User's Guide
one burst each time a bus trigger command is received. To trigger the function
generator from the remote interface (GPIB, USB, or LAN), send the *TRG
(trigger) command. The front-panel
generator is waiting for a bus trigger.
phase remain in effect but the burst period is ignored.
(equivalent to TRIG:SOUR IMM command).
(wait) command. When the *WAI command is executed, the function
generator waits for all pending operations to complete before executing any
additional commands. For example, the following command string guarantees
that the first trigger is accepted and the operation is executed before the
second trigger is recognized.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
(operation complete) command to signal when the burst is complete. The
*OPC? command returns "1" to the output buffer when the burst is complete.
The *OPC command sets the "Operation Complete" bit (bit 0) in the Standard
Event register when the burst is complete.
Remote Interface Reference
key is illuminated when the function
4
235