Repairing the High Pressure Flow Cell

When required
Tools required
Parts required
Preparations
1
Use a 4 mm hex key to unscrew the window assembly [1]
and remove the gasket [2] from the cell body.
1
2
1100 Series DAD and MWD User Manual
If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
For parts see
"High Pressure Flow Cell"
Turn the flow off.
Remove the front cover.
Remove the flow cell, see
If you want to replace the gasket only, continue with
step 7.
2
1
on page 160
"Exchanging a Flow Cell"
Maintenance
8
on page 121.