Patient Support Base Adjustment - Planmeca Viso Technical Manual

Hide thumbs Also See for Viso:
Table of Contents

Advertisement

5 Adjustments and calibrations

5.11 Patient support base adjustment

Attaching calibration tools
110
Planmeca Viso
2. Lift the sensor coupling up (2) and rotate the angle sensor axle (3) until
the green value appears on the display.
3. Reattach the sensor coupling. Make sure that the sensor coupling does
not touch the sensor body.
4. Perfrom the automatic position sensor calibration.
1. Remove the tube assembly covers and the lower cover of the C-arm as
well as the upper arm cover. Refer to section "Removing covers" on
page 122.
2. Remove the cover of the patient support table, refer to section
"Removing patient support covers" on page 130.
3. Remove tube head front covers, see section "Removing C-arm covers"
on page 124.
4. Attach the Ball phantom to the patient positioning mechanism adapter.
5. Enter the calibration mode: Touch the service spanner on the Main
display. Touch Technical and enter password (1701). Touch Calibration
(3200) and select Imaging Arm (3250).
6. Touch the Drive to pin point button.
7. Switch off the X-ray unit.