Intra Error Codes
ERROR CODE
Exposure key was released too early during the exposure.
E.00
- check the control panel cable, telephone cable, and the arm cable.
- replace the control panel or use another exposure switch.
X-ray tube Anode voltage (kV) overshoot.
E.10
- run an preheat cal
- replace tubehead pcb, tubehead, then generator board. (this order)
X-ray tube Anode voltage (kV) dropped suddenly.
E.11
- run an preheat cal
- replace tubehead pcb, tubehead, then generator board. (this order)
X-ray tube cathode filament preheating voltages are not calibrated.
E.12
- run an preheat cal
- replace tubehead pcb, tubehead, then generator board. (this order)
Filament preheating voltage calibration failed.
E.13
- do preheat again
- replace tubehead pcb, tubehead, feedback cable
Membrane keyboard key short-circuited/ pressed during self test or faulty display
board.
E.29/36/57
- press Select to see re-occurs, if so replace membrane.
- replace membrane then control panel (this order)
kV value does not reach or it exceeds the given value (difference more than 5%).
- check incoming mains voltage (must be
E.30/34
- run an preheat cal, check feedback cable, check arm cable.
- replace generator pcb, tubehead pcb, tubehead (this order)
X-ray tube Anode current (mA) missing, or not in specified limits.
E.31/33
- run preheat cal, check feedback cable, arm cable.
- replace tubehead pcb, tubehead (this order)
X-ray tube Filament voltage (V) missing, or outside the range (too low or too high).
E.33
- see above
X-ray tube Anode voltage (kV) missing, or below the specified limit.
E.34
- see above
Too long of an exposure.
E.36
- switch out control panel
- replace membrane, control panel, tubehead pcb. (this order)
kV feedback signal open circuit or short circuit.
E.37
- check arm cable, check feedback cable.flip feedback cable over.
- replace tubehead pcb, tubehead. (this order)
mA feedback signal open circuit or short circuit.
E.38
- check arm cable, check feedback cable. flip feedback cable over
- replace tubehead pcb, tubehead. (this order)
Quick Reference Manual
ERROR MESSAGE EXPLANATION
+/-
5% of 110)
12