3. Adjust the position of support bars with the adjustment screws (1-4). E.g.
loosening the screw (1) and tightening the screw (2) will move the the
right support bar outwards.
The rear head support should be centred in the patient camera image.
Adjust the rear head support and calibrate according to the directions on the
touch screen.
Enter calibration mode:
Touch the service spanner on the Main display.
Touch Technical and enter password (1701).
Touch Calibrations (3200) and select Rear head support (3220).