CC/EC Beam Alignment

CC:
PAN
1. Hold CTL and CEPH key for 6 seconds.
2. Press CTL once let go, and then press the Collimator Selection key let go.
3. Press Exposure button to check beam.
4. If the beam is too left/right, use the mm +/- to move the beam. (Figure 4)
NOTE:
The beam will not move while exposing. Adjust the beam and then press the exposure button to check
alignment.
5. Press the Collimator Selection key once to save changes.
NOTE:
Pressing the Clock key will also advance the collimator to next available beam position.
6. Press CTL then the Collimator Selection key to exit beam alignment mode.
7. Press the CEPH key to exit Service mode.
CEPH
1. Hold CEPH.
2. Follow the Pan instructions.
EC:
PAN
1. Hold CTL and Hidden key for 6 seconds to enter service mode.
2. Press CTL once let go, and then press the Clock key let go to enter beam alignment
mode.
3. Press Exposure button to check beam
4. If the beam is too left/right, use the mm +/- to move the beam
NOTE:
The beam will not move while exposing. Adjust the beam and then press the exposure button to check
alignment.
5. Press the Clock key once to save any changes.
NOTE:
Pressing the Clock key will also advance the collimator to next available beam position.
6. Press CTL then the Clock key to exit beam alignment mode
7. Press the Hidden key to exit Service mode.
CEPH
1. Hold P, press exposure +/- to select Ceph collimator 4, 5, or 6. Let go of P.
2. Follow Pan instructions.
Quick Reference Manual
18