3.4.2 Ventilator switch-on / " SELF TEST " phase

Set the main switch (placed on the back of the ventilator) to "I".
Make sure that on the ventilator keyboard
(commands area) the green led (that indicates the
presence of mains power supply) is on.
Hold the ON-OFF key for few seconds to start the ventilator.
3-14
" SELF TEST " phase. Close the patient circuit.
During " SELF TEST " phase, the ventilator software carries out the
self-diagnostic tests and checks a series of devices necessary for safe
ventilator operation.
The ventilator turns on and the
automatic " SELF TEST " phase
begins.
Turbine functioning check and
status.
The system pull out the O
excess in the device in order
to calibrate the 21% O
concentration in a proper way.