Preparation to use

6.3.8 Protection fuses

Protection fuses are foreseen in the following circuit:
Main power supply
Supplementary outlet (max. 6 A)
12Vdc led power supply
5Vdc power supply for valves group heating
5Vdc power supply for elettronics
12Vdc power supply for flowmeter box
12Vdc ventilator power supply
12Vdc battery power supply
WARNING! Risk of injury for the user / patient
The operations described here below must be carried out by highly
qualified technical personnel, trained and formally authorised by SIARE
only.
In case of break of protection fuse:
interrupt the mai power supply;
eliminate the abnormal or the cause which caused the
fuse break,
replace the protection fuse with one with the same value
and technical characteristics.
WARNING! Risk of injury for the user / patient
Fuses with wrong values and technical characteristics can compromise
equipment integrity and safety.