3.2.2 Protection fuses
There are protection fuses installed on the
following circuits.
220 VAC power supply: protection fuses of
the circuit ( 2 x 1AT ).
Battery supply: protection fuse of the battery
circuit (1 x 10AT).
Fuse replacing
If a protection fuse breaks, please proceed as follows.
Cut OFF mains power supply.
Remove the fault or the cause that caused the fuse breakage.
Replace the fuse with another one that has the same value and the same technical
characteristics.
Falco 202 Evo
WARNING !! Patient / clinician injury hazard
The operations described below must be carried out only by highly qualified
staff, specifically trained and authorised by SIARE.
WARNING !! Patient / clinician injury hazard
Using fuses of incorrect value or with incorrect technical features might
affect the lung ventilator's integrity and safety.
3-7