4.5.4 Protection fuses

There are protection fuses installed on the following
circuits.
220 VAC power supply: protection fuses of the
circuit ( 2 x 1AT ).
Battery supply: protection fuse of the battery circuit
(1 x 10AT).
WARNING !! Risk for Patient / User injury
The operations described here below must be carried out by highly qualified
technical personnel, trained and formally authorised by SIARE only.
In case of break of protection fuse, please proceed as follows.
Cut OFF the main power supply.
Remove the fault or the cause that caused the fuse breakage.
Replace the protection fuse with another one that has the same technical characteristics.
WARNING !! Risk for Patient / User injury
Fuses with wrong values and technical characteristics can compromise equipment
integrity and safety.
4-18
User manual, DU33ND101