Connection Of Anaesthetic Gases Scavenging System; Use Of Antibacterial Filter - Siare Morpheus ND User Manual

Anaesthesia unit
Table of Contents

Advertisement

4.3.4 Connection of anaesthetic gases scavenging system

SCAVENGINGS : gas scavenging connector.
O2 : gas supply connection for active gas scavenger.

4.3.5 Use of antibacterial filter

Apply the antibacterial filters to the patient circuit.
WARNING !! Risk of injury for the patient
To protect the patient from particles and dust, generated for example by the soda
lime, it is necessary to use a filter between the inpiratory hose and the patient, i.e.,
the filter on the Y connector or the filter on inspiratory hose.
WARNING !! Risk of injury for the patient
Replace the antibacterial filters as indicated in the maintenance instructions (please,
see on chapter 7).
Morpheus_ND
For more details about scavenging system manufactured by SIARE, please refer to
relative User and Service manual.
WARNING !! Risk for Patient injury.
If the scavenging collector of the anaesthesia unit is blocked, in the breathing
system and in patient lungs could create a negative pressure.
Always ensure that the collector is not obstructed and/or blocked.
4-9