Use of antibacterial filters

Use antibacterial filters on patient circuit.

3.3.5 Data Connection (Trend and Events downloading )

Falco 101
WARNING !! Patient injury hazard
To protect the patient from any dust and particles, you must install a filter
between the inspiratory tube of the respiratory circuit and the patient.
WARNING !! Patient injury hazard
Replace the antibacterial filters as indicated in the maintenance
instructions.
WARNING !! Patient injury hazard
Please perform the "SELF TEST" phase every time you replace the
patient circuit.
The system will check the patient circuit every time you turn on the
ventilator.
WARNING !! Strangulation hazard.
Pay utmost attention when connecting the patient to the ventilator.
If not carefully placed, the tubes, cables, the patient circuit and other
similar components installed on the ventilator might put the patient at
risk.
WARNING !! Burns hazard.
Do not use conductive masks or respiratory tubes during surgery with
electrosurgical unit: they might cause burns.
Connect the interface cable to the ODU connector (RS-232) and to a PC
serial port.
A dedicated software permits the download and the analysis of the data
stored in the lung ventilator.