Instrument Messages
Event Messages
Err#
Message
-200
Execution error; No
ranges are defined.
Activate a range.
-200
Execution Error;
Preselector centering
failed
-200
Execution Error;
Signal not stable
enough to track
-200
Execution Error; Store
ref trace before
turning on Normalize
-200
Execution error; Sync
word was not found.
-200
Execution error; Trace
file contains no
compatible traces.
-200
Execution error; Trace
file created by
incompatible version
of Phase Noise App
-200
Function not available
before Marker Zoom is
performed
-200
Function not available
before perform
Disturbance Zoom
-200
Function not available
before Signal Zoom is
performed
-200
Must perform Scan
before do Search
-200
No marked signal
-200
No Measure At Marker
Added to Signal List
Instrument Messages
Verbose/Correction Information
There are no active ranges in the range table. You will need to activate
at least one range.
Algorithm failed to center the preselector. This may be caused by the
signal peak being too low in amplitude, or it could be due to excessive
CW input signal, alignment error, or hardware failure.
The signal that you have selected to track is changing too much for the
function to track it properly.
The Reference trace data must be stored in the Ref trace before you
turn on the Normalization function.
NADC & PDC: In an EVM measurement, the sync word is not found and
the synchronization cannot be established when Sync Word is
selected in the Burst Sync menu.
Flexible Digital Demodulation: The sync word cannot be detected
because of inappropriate parameter settings or incorrect signal.
The trace file may have been created by another version of the Phase
Noise personality, which uses a different trace format that is
incompatible with the version you are running. Please check you are
running the most up to date version of the personality.
The trace file may have been created by another version of the Phase
Noise personality, which uses a different trace format that is
incompatible with the version you are running. Please check you are
running the most up to date version of the personality.
Function not available before perform Marker Zoom
Need to zoom in before you can zoom out
Function not available before Signal Zoom is performed.
Cannot perform Search as the trace data is found empty
Cannot perform the selected function because no signal was marked.
You must mark the peak of interest before selecting the function.
No signal peak was added into Signal List as there is no valid measure
to Marker result. You must perform Measure at maker before selecting
Measure At Marker --> List.