Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; MPA
GPS port power setting
<n>dBm is lower than
-130dBm minimum
-221
Settings conflict; MPA TX
port<n> amplitude
correction delta exceeds
<n>dB between port<n1>
and port<n2>
-221
Settings conflict; MPA TX
port<n> amplitude
correction value <n>dB is
out of range. The valid
range is <n> ~ <n>dB"
-221
Settings conflict; MPA TX
port<n> power setting
<n>dBm is lower than
-130dBm minimum
-221
Settings conflict; MS
gain is not available in
this Mode
-221
Settings conflict; Must
apply Amplitude
Correction to make this
unit available
-221
Settings conflict; No Cal
Factor for Power Sensor
-221
Settings conflict; No
meas frequencies are
above 3.6 GHz
-221
Settings conflict; no
source selected
-221
Settings conflict; No
Stepped Scan Type
-221
Settings conflict;
Normalize is not
available when Scale Type
= Lin
Instrument Messages
Verbose/Correction Information
Indicating the power output setting (UI power + Amp Corr Value)
on MPA GPS port is lower than supportable minimum -130dBm.
Set power output setting to be larger than -130dBm to eliminate
this warning.
Indicating the amplitude correction delta value between MPA TX
port<n1> and port<n2> for MXG mode (port difference) exceeds
max loss delta value.
Set related amplitude correction delta value to be lower than the
max loss delta value given in error message.
Indicating the amplitude correction value of MPA TX port<n> for
MXG mode is out of range.
Set related amplitude correction value within the valid range
given in error message.
Indicating the power output setting (UI power + Amp Corr Value)
on MPA TX port<n> is lower than supportable minimum
-130dBm.
Set power output setting to be larger than -130dBm to eliminate
this warning.
Mobile Station gain correction is not available in some Modes, or
in some measurements (for example, the SA measurement).
These special units only apply when you are doing antenna
measurements so you have to have a correction which includes
Antenna Units enabled
The center frequency set has no cal factor data associated with it.
Optimize Preselector can only be performed on frequencies in
high band i.e. freqs above 3.6 GHz. The current setup does not
have input freqs (IF) in this range so an Optimize Preselector
cannot be performed.
You must select a source using Select Source before you can do
this
Discrete Scan Type does not appear in this model
Normalize does not support Linear amplitude scale, since the
results are always presented as a dB ratio.