Settings Conflict; Freq > 3.6 Ghz Unavailable While Electronic Attenuator Enabled - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Freq >
3.6 GHz unavailable while
electronic attenuator
enabled
-221
Settings conflict;
frequency is outside
available range
-221
Settings conflict;
Function not available in
Correction Trace Display.
-221
Settings conflict;
Function not available in
External Mixing
-221
Settings conflict;
Function not available in
Tracking Source mode
-221
Settings conflict;
Function not available
while running multirange
scan
-221
Settings conflict;
Function not available
with current LISN Type
-221
Settings conflict;
Function not available
with time domain scan
type.
-221
Settings conflict;
Function only available
in Tracking Source mode
-221
Settings conflict;
Function unavailable with
MW Presel off
-221
Settings conflict;
Function unavailable with
this EMC Standard
-221
Settings conflict; Gate
control is Edge for Gated
FFT
-221
Settings conflict; Gate
control must be Edge for
this Gate Source
Instrument Messages
Verbose/Correction Information
The electronic attenuator does not function above 3.6 GHz. So if
you have that attenuator enabled, you cannot change the center
frequency so that frequencies above 3.6 GHz are
displayed/measured.
The desired frequency is not a valid setting.
The correction trace for the selected range is turned on. There is
no range to setup, If you want to edit the range, turn off
correction trace display.
The frequency offset feature cannot be used when you have
selected a log scale for the frequency axis.
The feature cannot be used when you are using a Tracking
Source.
Settings change is not allowed while multirange scan is running.
You must stop the current measurement if you wish to change the
settings.
Not all functionality is available for all LISN types
Scan Time and Points not supported in TDS
The feature cannot be used unless you are using a Tracking
Source.
You cannot center or adjust the preselector because the
Microwave Preselector is currently off
Filter BW is auto-coupled when EMC Standard is set to CISPR or
MIL. You must set the EMC standard to None if you are to change
the Filter BW.
You cannot use level triggering to control the gate if you are using
the FFT gating method.
You cannot use level triggering to control the gate when you are
using the currently selected gate source.